TB 5 5 - 4 9 2 0 - 4 2 9 - 3 5
11.
Resistance Test.
a.
Performance Check.
(1) Connect resistance standard (A4) leads (B4) and (B5)
with adapters (B1) and (B2) to the TI J1-H (HI) and J1-J (LO)
connector.
(2) Position TI controls (a) thru (e) as follows:
(a)
(c)
(d)
(e)
METER SOURCE switch to CONTINUITY.
(3) Set TI 115 vac power and 28 vdc power switches to ON.
Set resistance standard to "O" settings.
(4)
TI DMM will
indicate 0.000 0.002.
(5) Adjust resistance standard controls (A4) for 1,000
ohms.
TI DDM will indicate between .980K and 1.020K ohms.
20