TB 9-4920-367-35
NOTE
Unless otherwise specified, verify the result of each test and,
whenever the test requirement is not met, take corrective
action before continuing with the calibration. Adjustments
required to calibrate the TI are included in this procedure.
Additional maintenance information is contained in the
manufacturer's manual for this TI.
NOTE
Unless otherwise specified, all controls and control settings
refer to the TI.
7. Equipment Setup
NOTE
Insure that pneumatic pressure standard (A1) transducers
have been zeroed within the last 8 hours.
a. Remove brass fitting with nipple from flexible hose of TI.
b. Connect equipment as
shown in
figure 1.
(Connect TI to
0-20 PSIA inlet port on
pneumatic pressure standard.)
c. Open exhaust valve, shutoff valve, and inlet
valve on
pneumatic pressure controller
(B4).
d. Position controls on pneumatic pressure standard as indicated in (1) through (6)
below:
(1) Set POWER switch to ON and allow 1 minute for warm-up.
(2) SOURCE pushbutton to INT.
(3) UNITS DISPLAYED switch to IN HG.
(4) RANGE pushbutton to 0-20.
(5) SENSITIVITY pushbutton to LOW.
(6) Press RESET pushbutton.
e. Lightly tap TI and compare TI ambient atmospheric pressure indication to
indication on pneumatic pressure standard. If TI indication does not agree with pneumatic
pressure standard within 0.25 in. Hg, adjust the calibration adjustment (located beneath
the glass face of TI) to agree with pneumatic pressure standard indication.
4