TB 9-4931-383-50
Figure 3. Model 332A top view - adjustment locations.
(2) Repeat a above until no further adjustments are required.
10. Output Accuracy and Tracking
a. Performance Check
(1) Set voltage divider (A7) to one-tenth of actual value of dc transfer standard (A2)
(2) Set dc transfer standard for 1.018 + ∆E, mean output and set ∆E dials to last
three digits of standard cell value.
(3) Position TI controls as listed in (a) through (c) below:
(a) VOLTAGE RANGE switch to 10.
(b) Voltage dials to 10.000000.
(c) POWER switch to ON.
(4) Adjust TI voltage dials for a null indication on null detector (A4).
12