CALIBRATION PROCEDURE FOR SPECTRUM ANALYZER
CALIBRATION PROCEDURE FOR SPECTRUM ANALYZER - Continued
SECTION I. IDENTIFICATION AND DESCRIPTION
Table 1. Calibration Description - Continued
SECTION II. EQUIPMENT REQUIREMENTS
SECTION III. CALIBRATION PROCESS FOR SPECTRUM ANALYZER IF PLUG-IN HEWLETT-PACKARD, MODEL 8552( )
Equipment Setup
Equipment Setup - Continued
Figure 1. 30 MHz calibration signal and harmonics.
Calibrator Output Amplitude
Calibrator Output Amplitude - Continued
Figure 3. Model 8552B adjustment locations - bottom view.
Figure 4. Test instrument - board A6 adjustment locations.
Table 4. Bandwidth Accuracy
Amplitude Display Accuracy
Table 5. Amplitude Display Accuracy
Figure 5. Scan time accuracy - equipment setup.
Table 6. Scan Time Accuracy
Power Supply Check
SECTION IV. CALIBRATION PROCESS FOR SPECTRUM ANALYZER RF PLUG-IN HEWLETT-PACKARD, MODEL 8553( )
Scan Width Accuracy
Tuning Dial Accuracy and Range
Table 7. Tuning Dial Accuracy
Figure 6. Model 8553( ) - adjustment locations.
Frequency Response
Figure 7. Frequency response (10 to 110 MHz) - equipment setup.
Average Noise Level (Sensitivity)
Figure 8. Noise level limit indication.
Figure 9. Spurious response - equipment setup.
Local Oscillator Stability
Figure 10. Local oscillator stability.
SECTION V. CALIBRATION PROCESS FOR SPECTRUM ANALYZER, RF PLUG-IN HEWLETT-PACKARD, MODEL 8554( )
Scan Width Accuracy
Tuning Dial Accuracy
Table 8. Tuning Dial Accuracy
Figure 11. Test instrument - adjustment locations.
Figure 12. Frequency response - equipment setup.
Average Noise Level (Sensitivity)
Figure 13. Spurious response - equipment setup.
Local Oscillator Stability
Figure 14. Local oscillator stability.
SECTION VI. CALIBRATION PROCESS FOR SPECTRUM ANALYZER RF PLUG-IN HEWLETT-PACKARD, MODEL 8555( )
Equipment Setup
Dial Accuracy
Scan Accuracy
Noise Sidebands
Residual Responses
SECTION VII. CALIBRATION PROCESS FOR SPECTRUM ANALYZER LF PLUG-IN HEWLETT-PACKARD, MODEL 8556( )
Equipment Setup
Equipment Setup - Continued
Tracking Generator Amplitude
Figure 13. Adjustment locations.
Marker Accuracy
Scan Width Accuracy
Figure 14. Scan width accuracy display.
Center Frequency Accuracy
Figure 15. Frequency response - equipment setup.
Average Noise Level
Spurious Responses
Figure 16. Spurious responses - equipment setup.
Figure 17. Intermodulation distortion products.
Final Procedure