TB 9-6625-1965-24
TRIGGER
LEVEL
SET
B
(8) Adjust frequency
counter
to
stop
count
(approximately 2.5 V).
(9) Repeat (7) and (8) above as necessary until frequency counter starts when
calibrator is in OPERATE; stops when calibrator is in STANDBY, and A and B TRIG
indicator lights are blinking in unison when calibrator is in OPERATE. Leave calibrator
in STANDBY after establishing trigger level.
(10) S e t m A S E C m A s w i t c h t o m A S E C a n d R E A D m A S E C - E R A S E s w i t c h
t o ERASE.
NOTE
In the following steps do not leave calibrator in OPERATE for
more than 1 second as monitored on frequency counter.
(11) Momentarily set calibrator to OPERATE. TI will indicate 00.0 0.1 mASEC; if
not, perform b (1) through (4) below.
(12) Set READ mASEC-ERASE switch to READ mASEC.
(13) Momentarily set calibrator to OPERATE. If TI does not indicate same reading
2 digits, perform b (5) through
in mASEC as frequency counter indicates in time,
(16) below.
(14) Press TI RANGE 2000MA pushbutton and repeat steps 12 a (3), (4), (5), (9),
and (10) above.
(15) Momentarily set calibrator to OPERATE. TI will indicate 000
1 mASEC.
(16) Set READ mASEC-ERASE switch to READ mASEC.
(17) Momentarily set calibrator to OPERATE. TI will indicate the same reading in
mA as frequency counter indicates in time, 2 digits.
b. Adjustments
(1) Disconnect TI from equipment setup.
(2) Set READ mASEC-ERASE switch to ERASE.
(3) Set mASEC mA switch to mASEC. If TI does not indicate 0000, 1 digit, adjust
R14 ZERO (fig. 2) until TI indicates 0000.
(4) Using cables of equal length connect calibrator output to TI MA and COMMON
(5) Set mASEC mA switch to mA.
(6) Set calibrator for an initial 22 mA output and, if necessary, rotate calibrator
knob below EDIT FIELD pushbutton to adjust calibrator for a TI indication of 22 mA.
(7) Set calibrator to STANDBY.
(8) Connect digital multimeter to TI TP3 (fig. 2) and COMMON terminal.
(9) Set mASEC mA switch to mASEC.
10