TB 9-6625-2045-35
Figure 2. Ratio resistors check - equipment setup.
Table 8. Ratio Resistors
Test Instrument
CRL MULTIPLIER
indications
(MULTIPLIER)
CRL (CGRL)
(Ω)
switch
dial
Min
Max
10
99
101
Ω10 (Ω x 10)
10
990
1010
Ω100 (Ω x 100)
10
9900
10,100
kΩ1 (Ω x 1k)
10
99,000
101,000
kΩ10 (kΩ x 10)
2.5
247,500
252,500
kΩ100 (kΩ x 100)
0.8
792,000
808,000
MΩ11
1For
model 1650A only.
b. Adjustments. No adjustments can be made.
9