Table 17. Ac Voltage Accuracy.
Figure 9. AN/GRM-114A - Bottom View.

Calibration Procedure For Radio Test Set, An/Grm-114A (Ifr, Model Fm/Am-1100S) Page Navigation
  22    23    24    25    26  27  28    29  
img
. . . . . . . . .
TB 9-6625-2059-24
(26) Set TI multimeter RANGE switch to 0-10%.
(27) Adjust 1 kHz control until TI multimeter indicates 10 percent distortion. If
audio analyzer does not indicate between 8.5 and 11.5 percent distortion, perform b (8) and
(9) below.
(28) Set TI multimeter RANGE switch to 0-30%.
(29) Adjust 1 kHz control until TI multimeter indicates 30 percent distortion. If
audio analyzer does not indicate between 26.5 and 33.5 percent distortion, perform b (8)
and (9) below and adjust for best in-tolerance condition.
CAUTION
Do not exceed -30 dBm input to antenna.
(30) Set 1 kHz and VAR controls to OFF.
(31) Connect signal generator to measuring receiver input.
(32) Adjust signal generator frequency for 125.500 MHz and 60 percent AM (1 kHz)
modulation at -30 dBm.
(33) Connect signal generator to TI ANTENNA INPUT using attenuator (fixed). Set
AM-FM switch to AM and set TI multimeter RANGE switch to AM%. If TI multimeter
does not indicate between 50 and 70 percent, perform b (10) below.
b. Adjustments
(1) Adjust A2A1A1R65 (fig. 8) for 0 indication on multimeter.
(2) Adjust A2A1A1R73 (fig. 8) for 0 indication on multimeter.
(3) Adjust calibrator output for a 1.00000 indication. Adjust A2A1A1R69 (fig. 8) for
a 1V indication on TI multimeter.
(4) Adjust A2A1A1R119 (fig. 8) for ∞ (infinity) on TI multimeter.
(5) Set resistance standard to 30 Ω. Adjust A2A1A1R123 (fig. 8) for 30 Ω indication
on TI multimeter.
(6) Adjust calibrator output for a 1 V indication. Adjust A2A1A1R72 (fig. 8) for a 1
V indication on TI multimeter.
(7) Adjust A2A1A1R110 (fig. 8) for 6 dB indication on SINAD scale.
(8
Adjust 1 kHz control until audio analyzer indicates 10 percent distortion.
(9) Adjust A2A1A1R105 (fig. 8) for 10 percent indication on TI multimeter.
(10) Adjust A2A1A1R133 (fig. 8) for 60 percent indication on TI multimeter.