• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
SAMPLE CALIBRATION TEST REPORT

Calibration Procedure For Calibration Fixture (Notch Filter) Type 067-0938-00 Page Navigation
  4    5    6    7    8  9
img
TB 9-6625-2212-24
CALIBRATION MEASUREMENT VALUES
Frequency
Nominal value approximate
dB
(Hz)
(dBm)
Test
Initial test
Measured
instrument
Harmonic
Min
Max
report
value
100
k
2
10.0
11.0
100
k
3
6.5
7.5
100
k
4
5.0
6.0
100
k
5
4.0
5.0
50
k
2
9.5
10.5
50
k
3
6.0
7.0
50
k
4
4.5
5.5
50
k
5
3.5
4.5
10
k
2
9.0
10.0
10
k
3
5.5
6.5
10
k
4
4.0
5.0
10
k
5
3.0
4.0
1
k
2
9.0
10.0
1
k
3
5.5
6.5
1
k
4
4.0
5.0
1
k
5
3.0
4.0
20
2
9.0
10.0
20
3
5.5
6.5
20
4
4.0
5.0
20
5
3.0
4.0
DATE: ____________________________
Calibration Report Number: _________________
___________________________
Calibration Technician
___________________________
Facility Chief
A-3