TB 9-6625-2233-35
Test instrument
parameters
Performance specifications
Sine wave3
Output amplitude: <10 V p-p
Range: 2 mHz to 19.99 Hz
distortion
Accuracy: <1% (-40 dBc)
Range: 20 Hz to 99.9 kHz
Accuracy: <0.5% (-46 dBc)
Range: 100 to 999.9 kHz
Accuracy: <1 % (-40 dBc)
Range: 1 to 5.999 MHz
Accuracy: <2% (-34 dBc)
Range: 6 to 7.999 MHz
Accuracy: <2.5% (-32 dBc)
Range: 8 to 9.999 MHz
Accuracy: <3% (-30 dBc)
Range: 10 to 19.99 MHz
Accuracy: <4% (-28 dBc)
Output amplitude: >l0 V p-p
Range: 2 mHz to 19.99 Hz
Accuracy: <1 % (-40 dBc)
Range: 20 Hz to 99.9 kHz
Accuracy: <0.5% (-46 dBc)
Range: 100 to 999.9 kHz
Accuracy: <1 % (-40 dBc)
Range: 1 to 5.999 MHz
Accuracy: <2% (-34 dBc)
Range: 6 to 7.999 MHz
Accuracy: <3% (-30 dBc)
Range: 8 to 9.999 MHz
Accuracy: <4.5% (-27 dBc)
Range: 10 to 19.99 MHz
Accuracy: <5% (-26 dBc)
Sine wave4
Range: <1 MHz
Accuracy: <-60 dBc
Non-harmonically
related signals 200 Hz
Range: 1 to 9.99 MHz
of output
Accuracy: <-40 dBc
Range: 10 to 19.99 MHz
Accuracy: <-36 dBc
Square wave
Rise/falltime: <13 ns
<5%, 20 mV of p-p amplitude
Aberrations:
Symmetry: 10 to 90% in l% steps to l MHz linearly decreasing to 50%
fixed 20 MHz
Accuracy: (l% +20 ns)
See footnotes at end of table.
3