TB 9-6625-2246-50
(8) Set dc voltage divider dials to .999999TEN.
(9) Set calibrator output to dc reference standard 10 V output test report value.
(10) After reading has settled, operate sensitivity box L and H switches and null
meter RANGE switch while using calibrator output adjustment control to obtain best
null on null meter 10 V range.
(11) Record final calibrator output voltage indication as 10 V reference.
(12) Up-range null meter and set dc voltage divider dials to .1000000.
(13) Set calibrator output to 10 times dc reference standard 10 V output test
report value.
10