TB 9-6625-2316-35
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Connect adapter cable W100, supplied with TI, to J100 connector; then connect
electric connector assembly to the other end of adapter cable.
b. Connect power cable to the power connector and to a 115 ac power source.
c. Set POWER switch to A.C.
a. Performance Check
(1) Set the FUNCTION switch to CONT. P.
(2) Set the CABLE LENGTH switch to 250.
(3) Set AUTO/SINGLE switch to SINGLE.
(4) Press the RESET pushbutton.
(5) Set resistance standard No. 2 for 20.4Ω and connect between pins 1A and 1B
(fig. 1) of the
electric connector assembly (connected to
end of
adapter cable).
(6) Press the ADVANCE pushbutton once to test circuit 01. TI TEST light will be
on and the FAULT light will be on; if not, perform b below.
(7) Press the RESET pushbutton and set resistance standard No. 2 for 19.6Ω.
(8) Press the ADVANCE pushbutton once to test circuit 01. TI FAULT light will
remain off; if not, perform b below.
(9) Press the RESET pushbutton.
4