TB 11-6625-821-35
7. Preliminary Procedures
a. No preliminary procedures are required to prepare the unit under test for
calibration. However, procure the following parts to fabricate a cable necessary for
calibration:
Connector, MS 3106A-14S-5S
Clamp, AN 3057-6A
Wire, copper, insulated, stranded, No. 16 AWG (20 ft long)
Spaghetti 3/4 inch diameter (4 ft long)
b. Fabricate a cable for the BEARING INDICATOR in accordance with the details in
figure 1. This cable is
required to
connect the BEARING INDICATOR to
power and to
provide test points.
NOTE
When the performance check (a below) is within tolerance, do
not perform the corresponding adjustment (b below). When the
performance check is not within tolerance, perform the
corresponding adjustment before continuing with the
calibration procedure. When the performance check is not
within tolerance and no adjustment is specified, the deficiency
must be corrected before continuing with the procedure.
8. Static Power Inverter CV-2128/ARN-83 Output
a. Performance Procedure
(1) Voltage measurements.
(a) Connect the equipment as shown in figure 2.
Figure 2. Test equipment connections for voltage measurement of Static Power Inverter
CV-2128/ARN-83 output.
4