TB 9-4931-505-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Differential Comparator, Tektronix Type 5A13N. The manufacturer's manual
was used as the prime data source in compiling these instructions. The equipment being
calibrated will be referred to as the TI (test equipment) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 4
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Deflection factor
Range: 1 mV dc to 5 V dc in 12 calibrated steps of
1, 2, 5 sequence
3%
Accuracy:
Dc coupled
Dc to 2 MHz
Ac coupled
mode. Dc to 10 kHz within 2 kHz
Input R and C
Range: 1 M paralleled by about 51 pF; time constant
normalized for 51 s 3% between
channels
Maximum safe input voltage:
350 V (dc + peak ac, 1 kHz or less)
Dc input coupled
Ac input coupled
350 V dc
100 pA or less at +25 C
Maximum input gate current
Common-mode signal
Range: for deflection factor 1 to 50 mV/div, at least
15 V; for 0.1 to 5 V/div, at least 350 V
Range: 1 s to recover to within 3.0 V, and 0.1 ms to
Overdrive recovery
recover to within 1.5 mV after removal of
signal between 15V