TB 9-4931-535-35
(11) Repeat (3) through (7) above.
(12) Press keys and enter values using ENTRY keys as listed in (a) and (b) below:
(a) ENTRY CENTER FREQUENCY to 1 MHz.
(b) ENTRY CF STEP SIZE to 1 MHz.
(13) Repeat (3) through (7) above.
(14) Press keys and enter values using ENTRY keys as listed in (a) and (b) below:
(a) ENTRY CENTER FREQUENCY to 10 MHz.
(b) ENTRY CF STEP SIZE to 10 MHz.
(15) Repeat (3) through (6) above.
16. Local
Oscillator Sideband
a. Performance Check
(1) Disconnect rear panel OVEN REF OUT from EXT REF IN.
(2) Connect rear panel OVEN REF OUT to 50-75Ω input.
(3) Press keys and enter values using ENTRY keys as listed in (a) through (c)
below:
(a) ENTRY INSTR PRESET.
(b) ENTRY CENTER FREQUENCY to 10 MHz.
(c) MARKER/CONTINUOUS ENTRY COUNTER to on.
NOTE
Proceed when TI counter frequency indication is stable.
NOTE
60 Hz represents power line frequency; if power line frequency
is different than 60 Hz, use current power line frequency in
(4)(j) below.
(4) Press keys and enter values using ENTRY keys as listed in (a) through (l)
below:
(a) MARKER/CONTINUOUS ENTRY MKR → CF.
(b) MARKER/CONTINUOUS ENTRY COUNTER to off.
(c) MARKER/CONTINUOUS ENTRY OFFSET to on.