TB 9-6625-1014-35
(3) Adjust TI LRC dial, D-Q SERIES dial, and DET GAIN control for best null
indication on meter. D-Q SERIES - dial will indicate between 1.88 and 2.12.
(4) Repeat technique of (2) and (3) above, using resistance standard settings listed
in table 7. D-Q SERIES dial will indicate within limits specified.
Resistance standard settings
Test instrument D-Q SERIES dial indications
Min
Max
(Ω)
63.70
3.86
4.14
94.54
5.84
6.16
127.39
7.82
8.18
143.32
8.81
9.19
(5) Set FUNCTION switch to CDx0.1 SERIES.
(7) Adjust LRC dial D-Q SERIES dial, and DET GAIN control for best null
indication on meter. D-Q SERIES dial will indicate between 1.88 and 2.12.
(8) Repeat technique of (6) and (7) above, using resistance standard settings listed
in table 8. D-Q SERIES dial will indicate within limits specified.
b. Adjustments. No adjustments can be made.
15