TB 9-6625-106-24
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
FUNCTION GENERATOR 1 V dc
Agilent, Model 33250A
Accuracy: 1%
(33250A)
1 100 kHz @ 0.707 V rms
Accuracy: 1%
MEASURING RECEIVER
Spectral purity: measurement:
Measuring receiver system
Range: Carrier from 10 MHz to 50 GHz
N5530S consisting of:
Amplitude: +10 dBm to -60 dBm
Spectrum Analyzer, Agilent
Model E4448A (E4448A),
RF Level: Measurement:
Power meter, Agilent Model
Range: +10 to -95 dBm
N1911A (N1911A), and Sensor
Accuracy: 0.25 dB
module, Agilent Model
N5532A opt. 518 (518), 550
AM rate: measurement:
(550)
Range: 50%
Accuracy:
r1.25% at 1 kHz
r1.25% (1.5%) 3 to 26.5 GHz at 1 kHz
r1.25% (1.9%) 26.5 to 30 GHz at 1 kHz
FM deviation: measurement:
Range: FM .05 to 100 kHz
r2.5% 5 GHz at 300 kHz
r2.5% (r3.8% ) 20 GHz at 300 kHz
MICROWAVE
Range: 10 MHz
Anritsu, Model MF2414B003
(MF2414B003)
WORKSTATION
Range: 0 to 1 V dc
Fluke, Model 8846A
Accuracy: 1%
Range: 0 to 0.707 V rms @ 1 100 kHz
Accuracy: 1%
TIME/FREQUENCY
Frequency: 10 MHz
Datum, Model ET6000-75
Accuracy: 5 parts in 10-10 per day
WORKSTATION
(13589305)
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.