TB 9-6625-121-24
Output (Hz)
Min
Max
10
9.99
10.01
100
99.98
100.02
1000
999.8
1000.2
10 k1
9.998 k
10.002 k
19 k1
18.998 k
19.002 k
Increase input level as necessary to obtain stable readings at 10 and 19 kHz.
1
(5) Set function generator output to minimum.
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.