TB 9-6625-126-24
(16) Wait several minutes for indication on instrument controller screen to settle and
readjust if necessary.
(17) When indication on instrument controller screen is precisely 50.00 C (for R0)
and no further adjustment is necessary, press SET to accept new R0 setting.
Software
TI
Set-point
Sensitivity of
Range
Parameter
Effect
(C)
adjustment
2
50
RO
40 counts equals
Increase counts
about 0.1 C
increases temperature
2
200
ALPHA
40 counts equals
Increase counts
about 0.1 C
increases temperature
2
400
DELTA
2500 counts equals
Increase counts
about 0.1 C
decreases temperature
(18) Press and momentarily hold EXIT to return to return to temperature display.
(19) Repeat technique of (12) through (18) above for remaining set points and
parameters in table 6 except in step (14) (g), press SET one additional time to adjust Alpha
and two additional times for Delta.
(20) Return TI hot side to a safe temperature of approximately 50 C by repeating
step (12) above. Allow time for well to cool.
a. De-energize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.