• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
SECTION II. EQUIPMENT REQUIREMENTS
Duty Cycle

Calibration Procedure For Programmable Pulse/Function Generator, Model 8116A Page Navigation
  1    2  3  4    5    6    7    8  
img
. . . . . . . . .
TB 9-6625-127-24
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturers' manuals for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.
a. Press TI POWER button to turn TI on and allow 30 minute warm-up time.
b. Set TI controls as listed in (1) through (7) below:
MODE (Input) to NORM.
(1)
(2)
Function to square.
DTY to 50%.
(3)
CTRL to off.
(4)
AMP to 1 V.
(5)
OFS to 0 V.
(6)
COMPL (Output) to off.
(7)
8. Frequency
a. Performance Check
(1) Connect TI OUTPUT to frequency counter A input and set frequency counter for
50 Ω input.
(2) Set TI to first frequency listed in Applied column of table 3. If frequency counter
does not indicate within limits specified in table 3, perform b below.
(3) Repeat technique of (2) above for remaining frequencies listed in table 3. If
frequency counter does not indicate within limits specified in table 3, perform b below.
Table 3. Frequency
Test instrument
Frequency counter indication
Limits
Applied
Min
Max
50
MHz
47.500  MHz
52.500  MHz
10.0  MHz
9.500  MHz
10.500  MHz
1.0  MHz
0.950  MHz
1.050  MHz
10
kHz
9.700  kHz
10.300  kHz
1.0  kHz
0.970  kHz
1.030  kHz
1
1.00 Hz
0.9706 s
1.0312 s
100
mHz
9.681  s
10.341  s
Set frequency counter for period measurement.
1
4






Western Governors University

Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved. Design by Strategico.