TB 9-6625-1493-35
(13) Set calibrator to STANDBY.
(14) Repeat technique of (3) through (13) above for dc voltage divider settings and TI
RANGE switch positions listed in
table 11. Digital
multimeter/multimeter will
indicate
within limits specified.
b. Adjustments. No adjustments can be made.
Dc voltage
Test instrument
RANGE
divider
(V)
settings
switch settings
Min
Max
.010000
5 MV
0.49975
0.50025
.001000
500 V
0.49957
0.50042
.000100
50 V
0.49777
0.50223
24