TB 9-6625-170-24
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturers' manuals for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
(1) Plug TI into enclosure and turn on.
warm-up time.
8. Excitation Output, Load Regulation and Noise
a. Performance Check
(1) Connect Excitation Cable output between the respective amplifier channel (for
Amplifier #1 connect Excitation Cable to J1 input connection ect) and the multimeter hi
(2) Set the multimeter to read DC voltage.
(3) Set the EXCIT switch to 5V. Multimeter will read 4.95 to 5.05 V. If not,
perform b below. Place a 50 Ω resistor across the output. The output should not vary more
than 0.1%.