TB 9-6625-1957-24
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance applications which pertain
Test instrument
parameters
Performance specifications
Tektronix, Type DM 501
Dc voltage
Range: 0 to 1000 V (in 4 ranges)
Range:
Accuracy:
2 V, 20 V, 200 V, 1000 V
(0.1% of reading + 2 counts)
Ac voltage
Range: 2 V, 20 V, 200 V, 500 V
Accuracy:
20 Hz to 40 Hz
(1.2 % of reading + 2 counts)
40 Hz to 10 kHz
(0.7 % of reading + 2 counts)
10 kHz to 20 kHz
(1.2 % of reading + 2 counts)
Dc current
Range: 0 to 2 A (in 4 ranges)
Range:
Accuracy:
2 mA, 20 mA, 200 mA, 2 A
(0.2% of reading + 10 counts)
Ac current1
Range: 2 mA, 20 mA, 200 mA, 2 A
Accuracy:
40 Hz to 1 kHz
(0.6% of reading + 2 counts)
1 kHz to 10 kHz
(0.6% of reading + 10 counts)
Range: 0 to 20 M (in 5 ranges)
Range:
Accuracy:
2 k 20 k , 200 k , 2.0 M
(0.3 % of reading + 2 counts)
20 M
(0.5 % of reading + 2 counts)
Tektronix, Types DM 501A & DM 501AOPT02
Dc voltage
Range: 0 to 1000 V ( in 5 ranges)
Range:
Accuracy:
200 mV
(0.05 % of reading + 0.015 % of FS (3 counts))
2 V, 20 V, 200 V
(0.05 % of reading + 0.01 % of FS (2 counts))
1000 V
(0.05 % of reading + 0.02 % of FS (2 counts))
Ac voltage
Range: 0 to 500 V (in 5 ranges)
200 mV , 2 V, 20 V, 200 V
Accuracy:
20 Hz to 40 Hz
(1.0 % of reading + 0.05 % of FS (10 counts))
40 Hz to 10 kHz
(0.6 % of reading + 0.05 % of FS (10 counts))
10 to 20 kHz
(1.0 % of reading + 0.05 % of FS (10 counts))
500 V range:
Accuracy:
20 to 40 Hz
(1.0 % of reading + 0.2 % of FS (10 counts))
40 Hz to 10 kHz
(0.6 % of reading + 0.2 % of FS (10 counts))
10 kHz to 20 kHz
(1.0 % of reading + 0.2 % of FS (10 counts))