TB 9-6625-1965-35
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Ac voltage:
Fluke, Model 5720A 5700A/EP)
Range: 190 mV to 1000 V ac
(p/o MIS-35947), w/amplifier,
Accuracy: 0.157% for 45 Hz to 10 kHz 0.273 %
Fluke 5725A/AR (5725A/AR)
for 10 to 20 kHz
Dc voltage:
Range: 0 to 1000 V dc
Accuracy: 0.039%
Dc current:
Range: 0 to 2000 mA
Accuracy: 0.092%
Range: 10A
Accuracy: 0.15%
Range: 0 to 20 MΩ
Accuracy: 0.063%, 0.138% (20 MΩ range)
DIGITAL
Range: -17 to 1 V dc
Fluke, Model 8840A/AF-05/09
(AN/GSM-64D)
Range: 0 to 2000 ms
Fluke, Model PM6681/656
COUNTER
(PM6681/656)
Accuracy: 5 s
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturers' manuals for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.