TB 9-6625-2071-24
Table 1. Calibration Description Continued.
Test instrument parameters
Performance specifications
Range: 10, 100, 1000 V dc and auto ranging
Accuracy:
10, 100 V ranges: 0.03% of reading 0.004% of range
1000 V range: 0.087% of reading 0.004% of range
Remote trigger levels:
Range: -2 to +2 V
Option 010, AN/USM-459, and 5328AF096
Accuracy: 35 mV
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set, AN/GSM-286; AN/GSM-287; or AN/GSM-440. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parentheses.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
Table 2. Minimum Specifications of Equipment Required.
Manufacturer and model
Common name
Minimum use specifications
(part number)
Range: 105 to 125 V ac
Ridge, Model 9020A (9020A)
CALIBRATOR
Range: -9.5 to 950 V dc
Fluke, Model 5720A (5720A) (p/o
MIS-35947); w/ amplifier, Fluke
Accuracy: 0.0085%
5725A/AR (5725A/AR)
Range: 10 Hz to 1 MHz
(SG-1288/G)
Amplitude: 1 mV to 6 Vp-p
Range: -2 to +2 V dc
Fluke, Model 8846A (8846A/G)
Accuracy: 0.01%
Vertical: 500 mV to 6 Vp-p
Agilent, Model 54831M
Horizontal: 100 Hz to 20 kHz
(OS-303/G)
Accuracy: 3%
Range: 0 to 1 MΩ
IET Labs, HARS-LX-9-0.001
(HARS-LX-9-0.001)
Range: 5 to 512 MHz
Aeroflex, Model 2023B (2023B
Amplitude: 1 to 50 mVrms
OPT2/4/11/122)
Datum, Model ET6000-75
Range: 1 MHz
Accuracy: 1.25 parts in 10-10
(13589305)