TB 9-6625-2097-24
Paragraph
Page
SECTION
CALIBRATION PROCESS FOR RMS
CONVERTER, MODEL 13-4618-10
Preliminary instructions .......................................
Equipment setup .................................................
Converter output .................................................
Final procedure ...................................................
SECTION
CALIBRATION PROCESS FOR
FREQUENCY DEVIATION CONVERTER,
MODEL 13-4618-00
Preliminary instructions ....................................
Equipment setup .................................................
Span gain .............................................................
Final procedure ...................................................
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Recorders, Ink Writing, Gould, Model 2200S (Series); with RMS Converter,
Gould, Model 13-4618-10; and Frequency Deviation Converter, Gould, Model 13-4618-00.
The manufacturers' manuals were used as the prime data sources in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test instrument)
throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 3
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.