TB 9-6625-2163-35
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
FREQUENCY
Range: 1.025 to 1.099 GHz
Fluke, Model PM6681/656
COUNTER
Accuracy: .005%
(PM6681/656)
Range: 5 to 150 V dc
Fluke, Model 8840A/AF-05/09
(AN/GSM-64D)
Accuracy: 0.3%
PEAK POWER
Range: -5 to -21 dBm
Wavetek, Model 8502A
METER
(8502-16934-17071) w/power
Accuracy: 5%
(TEST SET, RADIO
detector, Wavetek Model 16934
FREQ POWER)
(16934)
PULSE
Range: 4 to 8 V
LeCroy, Model 9210 (9210)
w/plug-in, LeCroy, Model 9211
Pulse width: .5 to 25 s
(9211) (MIS 45839)
Trigger Delay: 65 s
RADAR TEST SET
Input:
(AN/UPM-155) w/accessory kit
Width Measurement: .025 s
Spacing Measurement: .025 s
Power Level: 0 to 10 dBm
Pulse Range: 0 to 3907 ms 3%
Output:
Accuracy: 0.01%
Pulse Source: 0 to 1 s
Power Level: 0 to 10 dBm
Timebase: .01 s to 4 ms .01%
Amplitude: 0 to 8 Vpp 3%
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in TM 11-6625-667-45.
d. Unless otherwise specified all controls and control settings refer to the TI.
CHANGE 1
4
PIN: 065659-001