TB 9-6625-2164-24
(19) Set CH 2 VOLTS/DIV switch to 20 mV, press AVE N, and adjust CH 2
POSITION to move display to center of screen.
(20) Adjust CH 2 POSITION control to move the two division display over the entire
graticule area. The CURSORS readout will not vary more than 2.4 mV.
(21) Remove all connections from TI.
a. Performance Check
(1) Press MENU/TEST and MEMORY DISPLAY 4, 5, 6 and MENU/TEST
pushbuttons.
(2) Set CH 1 VOLTS/DIV switch to 100 mV.
(3) Set TIME/DIV switch to 10 s.
t
(4) Connect function/arbitrary generator Output to CH 1 input, using 50
ermination, and press function/arbitrary generator Sine key.
(5) Set function/arbitrary generator frequency for 50 kHz and amplitude for 8
divisions on TI display.
(6) Set leveled sine frequency to 70 MHz.
(7) Set TIME/DIV to 50 ns. TI display will indicate at least 5.7 divisions.
(8) Move connection from TI CH 1 input to CH 2 input.
(9) Press AQR MODE CH 2 2 pushbutton and repeat (2) through (7) above for CH 2.
(10) Set TIME/DIV switch to 500 s.
(11) Set function/arbitrary generator frequency for 1 kHz and amplitude for 8
divisions centered on TI display.
(12) Set TIME/DIV switch to 50 ms.
(13) Set function/arbitrary generator frequency to 10 Hz. TI will indicate at least 5.7 divisions.
(14) Move connection from TI CH 2 input to CH 1 input.
(15) Press AQR MODE CH 1 1 pushbutton and repeat (10) through (13) above.
(16) Remove all connections from TI.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Press MENU/TEST and MEMORY DISPLAY 4, 5, 6 and MENU/TEST pushbuttons.
(2) Connect function/arbitrary generator Output to CH 1 and EXT TRIG inputs,
using BNC T and two 50 terminations.
(3) Set function/arbitrary generator frequency for 30 Hz and amplitude for 60 mV.
(4) Press TRIGGERING/MODE AUTO pushbutton.