TB 9-6625-2230-24
Test instrument
Calibrator
Calibrator error
range pushbutton
nominal
display indication
settings
settings
Min
Max
2 kΩ
1.000 kΩ
-0.09%
+0.09%
200 kΩ1
kΩ
-0.09%
+0.09%
100.00
MΩ
1.0000 MΩ
-0.17%
+0.17%
MΩ
MΩ
-0.5%
+0.5%
10.00
MΩ2
MΩ
MΩ
MΩ
50.0
49.2
50.8
MΩ3
MΩ
MΩ
MΩ
200
193
207
20 kΩ
10.000 kΩ
-0.09%
+0.09%
Set Calibrator 2 wire comp OFF.
1
2
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Set TI to measure frequency by selecting ac voltage mode and pressing HZ
pushbutton.
(2) Press 200 mV pushbutton.
(3) Adjust calibrator for output settings listed in table 7.
(4) TI will indicate within limits listed.
Test
instrument
Calibrator
Test instrument indications
range
pushbutton
Voltage
Indications
settings
settings
settings
Min
Max
200 mV
20.00
mV
100
Hz
99.94 Hz
100.06
Hz
200 mV
200.00
mV
19.000 kHz
18.990 kHz
19.010 kHz
200 mV
200.00
mV
190.00 kHz
189.90 kHz
190.10
kHz
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.