TB 9-6625-2277-35
Table 1. Calibration Description Continued
Performance specifications
Test instrument
(% of reading + digits)
parameters
(3 1/2 digit display)
Range: 200 A, 2 mA, 20 mA, 200 mA, 2000 mA
Dc current
Accuracy: (0.3 +1)
High current2
Range: 10 A
Accuracy: (0.5 + 1)
Range: 200 A, 2 mA, 20 mA, 200 mA
Ac current3
Accuracy: 45Hz to l0 kHz (1.0 + 2)
10 to 20 kHz
(2.0 + 2)
Range: 2000 mA
Accuracy: 45 Hz to 2 kHz
(1.0 + 2)
45 Hz to 3kHz
(1.0 + 2)1
High current2, 3
Range: 10 A
Accuracy: 45 Hz to 2 kHz (1.0 + 2)
Range: 200 Ω, 2 kΩ, 20 kΩ, 200 Ω
Accuracy: (.2 + 1)
Range: 2000 Ω and 20 MΩ
Accuracy: (.5 + 1)
Range: 2000 kΩ1
Accuracy: (.2 + 1)
Range: 2 Ω
Accuracy: (1 + 2)
Range: 20 Ω
Accuracy: (.5 + 2)
Range: 2 mS and 20 S
Accuracy: (.2 + 1)
Range: 200 nS
Accuracy: (1 + 10)
1Model
8010M only.
2Models
8010A and 8010A-01.
3Ac current and high current are verified during dc current checks since same shunts are utilized for both ac and dc current
functions.
4Model 8012A only.
REQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287 and AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2