TB 9-6625-2288-24
(43) Adjust calibrator for 2 V ac output. Adjust 3V MID (fig. 1) for a meter indication
between 1.85 and 2.15 V ac (ideal 2) (R).
(44) Adjust calibrator for 1 V ac output. Adjust 3V LO (fig. 1) for a meter indication
between 0.850 and 1.15 V ac (ideal 1) (R).
(45) Repeat (42) through (44) above until no further adjustments are required.
(46) Disconnect TI RF probe from calibrator.
a. Performance Check
(1) C o n n e c t c a l i b r a t o r W I D E B A N D O U T P U T ( w i t h o u t 5 0 o h m p a d ) t o T I
R F probe using 50 ohm feed through termination supplied with TI.
(2) Set RANGE-FULL-SCALE switch to .1-VOLTS.
(4) Adjust calibrator output voltage controls for 0.1 V indication on TI meter.
Calibrator voltage output will be within the limits specified in table 4.
(5) Repeat (3) and (4) above for remaining calibrator frequency settings listed in
Calibrator
Calibrator output voltage
(mV)
(kHz)
Min
(Max)
20
95
105
50
95
105
100
95
105
500
95
105
1000
95
105
Figure 2. Frequency response - equipment setup.