TB 9-6625-2297-24
Figure 1. Adjustment locations - test instrument bottom view.
11. Power Reference Level
a. Performance Check
(1) Press DATA ENTRY PRESET/LOCAL key then DATA ENTRY ENTER key.
NOTE
Power meter is a standard used to calibrate the TI.
Instructions in this text which refer to power meter setup or
connections refer to the standard power meter.
(3) Set power meter LINE switch to OFF position.
(4) Connect multimeter between center conductor of power meter rear panel VRF
connector and pin 1 of thermistor mount end of power meter interconnect cable.
(5) Round off multimeter indication to two decimal places and record value as R
(approximately 200 ).
(6) Disconnect multimeter from power meter interconnect cable.