TB 9-6625-2303-35
an 80 W indication on TI wattmeter. RF power meter will indicate between 75 and 85 W.
b. Adjustments. No adjustments can be made.
13. Plug-in Element
4410-2
a. Performance Check
(2) Insert TI into wattmeter and rotate TI until the arrow points in the direction of
50 Ω termination.
(3) Set factor switch to 1 position and allow equipment to warm-up for 5 minutes.
RF power meter will indicate between 75 and 85 W.
(6) Adjust signal generator frequency controls to 1.0 MHz and amplitude controls
for an 80 W indication on TI wattmeter. RF power meter will indicate between 75 and 85 W.
an 80 W indication on TI wattmeter. RF power meter will indicate between 75 and 85 W.
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA Label/Form in accordance with TB 750-25.