SECTION III. CALIBRATION PROCESS
Ac Voltage and Frequency
Calibration Procedure For A31U12661-3, Self-Alignment Interface Connection Device
Page Navigation
1
2
3
4
5
6
7
TB
9-6625-2304-24
Figure
1.
Dc
and
ac
voltage.
Table
3.
Dc
Voltage
Multimeter
indications
INPUT HI
(V)
to
test
instrument
Min
Max
TP25
0.89550
0.90450
TP24
8.95500
9.04500
TP22
17.9100
18.0900
b.
Adjustments.
Ensure
that dc
power
supply
output
is 36
(1.0)
V
dc;
otherwise,
no
adjustments
can
be
made.
5