TB 9-6625-2321-35
(5) Click on Settings.
(7) Enter 1.000 in Current Shunt Resistor window.
(8) Set calibrator output and TI controls to values listed in table 6.
Test instrument
Calibrator
range pushbutton
output
Test instrument indications
settings
settings
(mA)
(mA)
(mA)
Min
Max
20
19.0
18.800
19.200
200
190
188.00
192.00
b. Adjustments. No adjustments can be made.
a. Performance Check
Set calibrator output and TI controls to values listed in table 7.
Calibrator
output
Test instrument
Test instrument indications
settings
range pushbutton
(mA)
@ 1000 Hz
settings
(mA)
(mA)
Min
Max
20
19.0
18.400
19.600
200
190
184.00
196.00
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.