TB 9-6625-2328-35
(g) CHANNEL 2 Trigger/Sensitivity until AUTO TRG: is displayed, then
⇒ to display AUTO TRG: OFF.
(h) CHANNEL 2 Trigger/Sensitivity until LEVEL: is displayed, use ⇒ , ⇑ ,
⇐ , ⇓ to display LEVEL: .500 V and then Enter.
(i) CHANNEL 2 Trigger/Sensitivity until SLOPE: is displayed, then ⇒ to
display SLOPE: NEG.
(j) CHANNEL 2 50Ω /1MΩ to 50Ω (LED on).
(k) CHANNEL 2 DC/AC to DC (LED on).
(6) Set pulse generator output as listed in (a) through (k) below:
(a) Period to 10.00s.
(b)
Width to 150.0ns .
(c)
Vhigh to 1.000V.
(d)
Vlow to 0mV.
(e)
Delay to 0.00ns .
(f)
Lead to 1.00ns .
(g)
Trail to 1.00ns .
(h)
2 Pulse to OFF.
(i)
Loadcomp to OFF.
(j)
Trigger Mode to Normal.
(k)
Enable output.
(7) Press TI Stop/Single key. If TI does not indicate between 0.146 and 0.154 s,
perform b below.
b. Adjustments
(2) Press and release TI POWER ON/STBY pushbutton to STBY.
(3) Press and hold MATH Scale & Offset key while pressing POWER ON/STBY
pushbutton to ON until CAL SECURE is displayed.
(4) Press MATH Scale & Offset key until CODE:
0 is displayed.
(5) Press ⇐ key until CODE: 00000 is displayed.
(6) Press ⇒ , ⇓ , ⇐ , ⇑ keys until CODE: 53131 is displayed then press Enter key
and verify CAL UNSECURE is displayed.
(7) Connect function
generator
Function
Outputs
Unbalanced to
TI
CHANNEL 1 input.
5