TB 9-6625-2337-24
Table 30. Video Reset Output Characteristics - Continued
TI
Oscilloscope
Test description
Settings
Measurement type
Minimum
Maximum
RESETS 1 8 μs
----------------------------------------
Width
7.8 μs
8.2 μs
PW
RESETS 4.5 V Lvl
----------------------------------------
Top
3.5 V
5.5 V
1.75 s
2.25 s
Width
RESETS 2,
Use the arrow keys to highlight
2 s PW
RESET PULSE 1:
ENTR (OFF)
Use the arrow keys to highlight
RESET PULSE 2:
ENTR (ON)
4.25 s
3.75 s
RESETS 4,
Use the arrow keys to highlight
Width
2 s PW
RESET PULSE 2:
ENTR (OFF)
Use the arrow keys to highlight
RESET PULSE 3:
ENTR (ON)
14.9 s
15.1 s
1 TO 3 DELAY
Use the arrow keys to highlight
Delta time
RESET PULSE 1:
ENTR (ON)
9.9 s
10.1 s
Delta time
1 TO 2 DELAY
Use the arrow keys to highlight
RESET PULSE 3:
ENTR (OFF)
Use the arrow keys to highlight
RESET PULSE 2:
ENTR (ON)
26.0 s
24.0 s
Delta time
Use the arrow keys to highlight
1
PULSE 1 DELAY
PUSLE 1 DELAY:
Channel 1 to
25 s
25, ENTR (25 s)
channel 2 pulse 1
3499.0 s
3501.0 s
Delta time
Use the arrow keys to highlight
PULSE 1 DELAY
3500 s
PUSLE 1 DELAY:
Channel 1 to
3500, ENTR (3500 s)
channel 2 pulse 1
1Connect
TRIGGERS 0 OUT to oscilloscope channel 1 using 75
feedthrough termination.
(6) Reduce all outputs to minimum and disconnect equipment setup.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect TI RADAR SIM NORTH TRIGGER OUT to oscilloscope channel 1
using a 75 feedthrough termination.
(2) Connect TI RADAR SIM POS READOUT GATE to the oscilloscope channel 2
input using a 75 feedthrough termination.
(3) Press TI keys as listed in (a) through (s) below:
Press FUNC, ENTR keys.
(a)
Use the arrow keys to highlight GENERAL MENUS and press ENTR key.
(b)
14, ENTR (PRF/PRI).
(c)
Use the arrow keys to highlight PRF:.
(d)
300, ENTR (300PPS).
(e)