TB 9-6625-2337-35
TI
Oscilloscope
Minimum
Maximum
(s)
(s)
Test description
Settings
M1 SIF 0 s DLY DIF
---------------------------------------
-0.05
0.05
M1 SIF 50 s DLY
Press FUNC, ↑
49.95
50.05
5, ENTR (TIMING, VIDEO LEVELS)
Use the arrow keys to highlight 1ST REPLY
VIDEO:
5, ENTR (5.0V)
Use the arrow keys to highlight 2ND REPLY
VIDEO:
5, ENTR (5.0V)
Press FUNC, ↑
4, ENTR (TIMING, VIDEO LEVELS)
Use the arrow keys to highlight DELAY:
50, ENTR (50.00 s)
M4-1 0 s DLY DIF
-0.05
0.05
Use the arrow keys to highlight DELAY:
0, ENTR (0.00 s)
Use the arrow keys to highlight REPLY SIGNAL:
ENTR (MODE 4-1)
Press FUNC, ↑
3, ENTR (1ST REPLY)
Use the arrow keys to highlight REPLY SIGNAL:
ENTR (MODE 4-1)
Press FUNC, ↑
2, ENTR (CHALLENGES)
Use the arrow keys to highlight M1:
ENTR (OFF)
Use the arrow keys to highlight M4:
ENTR (SYNC)
M4-1 50 s DLY DIF
Press FUNC, ↑
49.95
50.05
4, ENTR (TIMING, VIDEO LEVELS)
Use the arrow keys to highlight DELAY:
50, ENTR (50.00 s)
(16) Reduce all outputs to minimum and disconnect equipment setup.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect the SUPPR OUT to channel 1 of the oscilloscope using a 75 Ω
feedthrough termination.
(2) Connect the IFF VIDEO CHAL/TAG to channel 2 of the oscilloscope using
a 75 Ω feedthrough termination.
(3) Press TI FUNC, ENTR keys and wait for TI to complete reset.
(4) Set oscilloscope to trigger on channel 1 and channel 1 measurement type to
Rise Time.
(5) Oscilloscope will indicate within limits listed in first row of table 27.