TB 9-6625-2339-35
Paragraph
Page
Resolution Bandwidth Switching Uncertainty .....
Absolute Amplitude Accuracy .............................
Resolution Bandwidth Accuracy .........................
Frequency Response.............................................
Displayed Average Noise .....................................
Residual Responses ..............................................
Power Supply........................................................
Final Procedure ....................................................
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Spectrum Analyzer, AN/USM-677 (Agilent, Model E4407B). The procurement
specifications and manufacturer's manual were used as the prime data sources in compiling
these instructions. The equipment being calibrated will be referred to as the TI (test
instrument) throughout this bulletin.
a. Model Variations. None
b. Time and Technique. The time required for this calibration is approximately 9
hours, using the dc and low frequency and microwave technique.
2. Forms,
Records, and
Reports
a. Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R)
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.