TB 9-6625-2340-24
(16) Amplitude indicated on TI display will be within limits specified in the first row
(17) Press TI keys as listed in (a) and (b) below:
(a)
.
(b)
(18) Repeat technique used in (13) (c) and (14) through (17) above for remaining TI
Table 28. Residual Response (5.7 GHz to 11.65 GHz)
Test instrument
Center frequency
Displayed amplitude
(GHz)
limit (dBm)
95
6.2
95
7.19
95
8.18
95
9.17
95
10.16
95
11.15
(19) Press TI keys as listed in (a) through (e) below:
(a)
(b)
VIDEO BW, 3, kHz.
(c)
TRACE A CLEAR WRITE.
(d)
SWEEP SINGLE.
(e)
(20) Wait for completion of sweep.
(21) Press TI MARKER PEAK SEARCH key.
(22) Amplitude indicated on TI display will be within limits specified in table 29.
Table 29. Residual Response (11.6 GHz to 12.5 GHz)
Test Instrument
Displayed
Start
Stop
amplitude limit
frequency
frequency
(dBm)
95
11.6 GHz
12.5 GHz
(23) Press TI keys as listed in (a) through (g) below:
(a)
(b)
RES BW, 100, kHz.
(c)
VIDEO BW, 3, kHz.
(d)
DISPLAY LINE ENTER, 85, dBm.
(e)
TRACE A CLEAR WRITE.
(f)
SWEEP SINGLE.
(g)
(24) Wait for completion of sweep.
(25) Press TI MARKER PEAK SEARCH key.
58