TB 9-6625-2340-35
(15) Repeat technique of (11)(c), (12) and (14) above for remaining TI REFERENCE
LEVEL and
synthesizer/level generator settings in
table12.
Synthesizer/level
Test instrument
Deviation marker ∆
Amplitude
Reference level
amplitude indication
(dBm)
(dBm)
(dB)
-3.9
-1.9
-5.9
-3.9
-7.9
-5.9
-9.9
-7.9
-11.9
-9.9
(16) Press TI MARKER OFF key.
(17) Press TI REFERENCE LEVEL key to select 0 dB.
(18) Change the synthesizer/level generator output level to -2.00 dBm and amplitude
increments to 0.1 dB steps.
(19) Press TI MARKER PEAK SEARCH and ∆ keys.
(20) Read the amplitude deviation from TI marker ∆ readout. Record displayed
amplitude level in Deviation Marker ∆ Amplitude Indication column of table 13.
(21) Repeat technique of (16) through (20) above for remaining TI REFERENCE
LEVEL key and
synthesizer/level generator settings in
table13.
Synthesizer/level
Test instrument
Deviation marker ∆
Amplitude
Reference level
amplitude indication
(dBm)
(dBm)
(dB)
0.0
-2.00
-2.10
-0.1
-2.20
-0.2
-2.30
-0.3
-2.40
-0.4
-2.50
-0.5
-2.60
-0.6
-2.70
-0.7
-2.80
-0.8
-2.90
-0.9
-3.00
-1.0
-3.10
-1.1
-3.20
-1.2
-3.30
-1.3
40