TB 9-6625-2342-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
the prime data sources in compiling these instructions. The equipment being calibrated
will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 5
hours, using the dc and low frequency technique.
2. Forms,
Records, and
Reports
a.
Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
parameters
Performance specifications
Vertical
Range: 2 mV to 5 V/div
Deflection &
Cursor accuracy: 2%
Volts cursors
Range:
DC to 150 MHz
Triggering
Range:
1 div @ 150 MHz
Internal A
2 div @ 150 MHz
Internal B
50 mV @ 150 MHz
External A
100 mV @ 150 MHz
External B
Horizontal
Range: 5 ns to 0.5 s/div
Sweep (A&B)
Accuracy: 1%
Range: 5 div from CRT center
Time cursors
Accuracy: 1%
Calibrator
Range: 0.4 Vp-p
Accuracy: 1%