TB 9-6625-2347-35
(9) SEC/DIV switches to .5 ms.
(10) SEC/DIV variable fully cw to detent and pushed in.
(11) VAR HOLDOFF fully ccw to NORM.
(12) TRIGGER MODE AUTO NORM TV FIELD switch to AUTO.
(13) SLOPE switch to
.
(14) INT switch to VERT MODE.
(15) SOURCE switch to INT.
d. Press TI POWER pushbutton to ON and allow at least 20 minutes for warm-up.
e. Adjust AUTO INTENSITY and AUTO FOCUS controls for suitable viewing.
a. Performance Check
(1) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 to TI CH 1 OR
X input and oscilloscope calibrator SOURCE/MEASURE CHAN 2 to TI CH 2 OR Y input.
(2) Set TI CH 1 VOLTS/DIV switch to 2m.
(3) Set oscilloscope calibrator CH 1 for a VOLTAGE output of 10 mV at 1 kHz.
(4) Adjust TI TRIGGER LEVEL and POSITION controls, as necessary, to view
(5) Rotate oscilloscope calibrator knob located below EDIT FIELD key for 5
divisions of vertical display. If oscilloscope calibrator Err display does not indicate within
limits specified in first row of table 3, perform b below.
(6) Repeat technique of (2) through (5) above for settings listed in table 3. If
oscilloscope calibrator err display does not indicate within limits specified in table 3,
perform b below.
Test instrument
Oscilloscope calibrator
VOLTS/DIV
Divisions of vertical
VOLTAGE
Err display
setting
deflection
output
Indication (%)
2m
5
10 mV
3
5m
4
20 mV
3
10 m
5
50 mV
3
20 m
5
.1 V
3
50 m
4
.2 V
3
.1 V
5
.5 V
3
.2 V
5
1V
3
.5 V
4
2V
3
1V
5
5V
3
2V
5
10 V
3
5V
4
20 V
3
10 V
5
50 V
3