TB9-6625-2350-35
Table 2. Minimum Specifications of
Equipment Required
Common name
Minimum use specifications
Manufacturer and model
(part number)
John Fluke, Model 5720A/CT (p/o
CALIBRATOR
Dc voltage:
MIS-35947), w/power amplifier,
Range: 0 to 1000 V
Accuracy: 0.0125%
John Fluke, Model 5725A/CT
(5725A/CT)
Dc current:
Range: 200 A to 5 A
Accuracy: 0.05%
Ac voltage:
Range: 0 to 1000 V
Accuracy: 0.125%
Ac current:
Range: 200 A to 5 A
Accuracy: 0.5%
Range: 190Ω to 100 MΩ
Accuracy: 0.0125%
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturers' manuals for this TI.
d. Unless otherwise specified, all control and control settings refer to the TI.
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.