TB 9-6625-3628-24
Table 2. Minimum Specifications of Equipment Required.
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATION ARM 6"
Range: 6 0.003 in.
Dynetics Inc.,
1 SEE TEST REPORT
Accuracy:
Model TCS1500-SARM
CALIBRATION ARM 48"
Range: 48 0.024 in.
Dynetics Inc.,
Accuracy:1 SEE TEST REPORT
Model TCS1500-LARM
Range: 0 to 10 V dc
Agilent, Model 3458A
Accuracy:1 0.01% of reading
(3458A)
Datum, Model ET6000-75
WORKSTATION
(13589305)
WEIGHT SET
Range: 0-375 lbs
Henry Troemner,
Accuracy:1 Class 6
Model 9475-SET
1
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The required peculiar accessories are listed in table 3.
Common name
Manufacturer (part number)
GEAR SHAFT 1
Dynetics Inc., (TCS1000-B0005)
p/o TCS1000
1
HOLDING FIXTURE
Dynetics Inc., (TCS1000-H0003)
p/o TCS1000
INSTRUMENT CONTROLLER
Hewlett Packard, (2710P)
INTERFACE BOX
Dynetics Inc., (TCS1000-Z0001)
p/o TCS1500-CALFIX
INTERFACE BOX METER CABLE
Dynetics Inc., (TCS1000-C0010)
p/o TCS1500-CALFIX
INTERFACE BOX SENSOR CABLE
Dynetics Inc., (TCS1000-C0011)
p/o TCS1500-CALFIX
JACK SET
Dynetics Inc., (TCS1000-J0001)
MACHINE KEYS1
Dynetics Inc., (TCS1000-K0001)
p/o TCS1000
SOFTWARE
DynCAL Torque Sensor Calibrator
TEST FIXTURE 1
Dynetics Inc., (TCS1000-H0002)
p/o TCS1000
TORQUE CELL CALIBRATION FIXTURE
Dynetics Inc., (TCS1500-CALFIX)
Supplied with TI, obtained from TCS1000 or procured through supply system (see figure 1).
1
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.