TB 9-6670-254-50
(10) Prepare a chart showing deviation of TI indication and standard weight and
affix to TI.
(11) Repeat (4) through (7) above for each cardinal point to be checked.
NOTE
The balance can only be used as a calibration standard for
b. Adjustments
(1) Adjust one of the three trimming nuts (fig. 1), which move horizontally on beam,
for an indication of 75 25 mV on digital voltmeter (R).
(2) Adjust either or both of the two trimming nuts (fig. 1), which move vertically on
the beam, for an indication of 950 150 mV on digital voltmeter (R).
18