TB 9-6695-273-35
(5) Repeat a (1) through (5) and if required adjust C1064 and C1066 (fig. 1) for best
in-tolerance condition (R).
(6) Repeat a (1) through (5) above as required due to interaction.
9. Signal Source Frequency
a. Performance Check
CAUTION
Do not connect TI directly to frequency counter. Damage to
frequency counter may result
(1) Connect 500 MHz +30 dBm output to frequency counter input C using (30 db)
attenuator.
(2) Press STANDBY pushbutton to ON.