TB 9-6695-293-50
Test instrument
parameters
Performance specifications
Calibrator, John Fluke, Model 5700A/( ) and 5720A/( )
Ac voltage (cont.)
100
300 kHz
.13 % + 15
22
mV
300
500 kHz
.18 % + 30
500 kHz 1.0 MHz
.48 % + 40
220
mV
10
20 Hz
600 + 16
20
40 Hz
240 + 10
40 Hz 20 kHz
110 + 10
20
50 kHz
360 + 10
50
100 kHz
900 + 30
100
300 kHz
.11 % + 30
300
500 kHz
.18 % + 40
500 kHz 1.0 MHz
.36 % + 100
2.2
V
10
20 Hz
600 + 100
20
40 Hz
180 + 30
40 Hz 20 kHz
85
+7
20
50 kHz
140 + 20
50
100 kHz
280 + 80
100
300 kHz
480 + 150
300
500 kHz
.12 % + 400
500 kHz 1.0 MHz
.24 % + 1.0 mV
22
V
10
20 Hz
600 + 1.0 mV
20
40 Hz
180 + 300
40 Hz 20 kHz
85
+ 70
20
50 kHz
140 + 200
50
100 kHz
280 + 400
100
300 kHz
600 + 1.7 mV
300
500 kHz
.14 % + 5.0 mV
500 kHz 1.0 MHz
.30 % + 9.0 mV
220
V
10
20 Hz
600 + 10 mV
20
40 Hz
180 + 3.0 mV
40 Hz 20 kHz
90
+ 1.0 mV
20
50 kHz
250 + 4.0 mV
50
100 kHz
600 + 10 mV
100
220 kHz
.16 % + 110 mV
1100
V
50 Hz 1.0 kHz
90
+ 4.0 mV
Range: 10 Hz to 1 MHz
Accuracy: 0.01%
Amplitude flatness, 1 kHz reference: (% of output + V)
Wideband ac voltage
(Option 03)
Voltage range
Frequency
1.1 mV
3 mV
>3 mV
10 Hz to 30 Hz
.3 + 0
.3 + 0
.3 + 0
30 Hz to 120 kHz
.1 + 0
.1 + 0
.1 + 0
120 kHz to
2.0 MHz
.2 + 3
.1 + 3
.1 + 3
2.0 MHz to 10 MHz
.4 + 3
.3 + 3
.2 + 3
10 MHz to 20 MHz
.6 + 3
.5 + 3
.4 + 3
20 MHz to 30 MHz
1.5 +
1.5 + 3
1.0 + 3
15
See footnotes at end of table.
5