TB 11-6625-2974-30
1-2.
Calibration Description. Table 1-1 lists the parameters to be
calibrated and the performance
specifications for each parameter.
Test Instrument
Performance
Parameters
Specifications
Regulator Filter A6AI
Dc output levels
+ 15.0 0.3 V
+ 10.0 0.3 V
-15.0 + 0.3V
+ 5.2 Volt Regulator A3A4
Dc output level
+5.20 0.05 V
Detector/Amplifier A3A2
Gain adjustment
0.0 + 0.1 Vdc
DVM Logic Module A5AI
ANT DETECTION Signal
Calibrate at 0.0 0.1 Vdc
Front panel METER display
Calibrate at O V and + 5 V
BITE Module A5A3
SYNTH LOCK signal threshold
0.4 0.1 Vdc above SYNTH output
ECCM Module A5A4
MT video threshold
-1.2 0.1 V
ECCM pulse width
500 5OnS, 1100 50 nS
Power Supply A1A1
Dc output levels
+ 10.0 0.4 V
+ 15.0 0.6 V (3 outputs)
-15.0 0.6 V (2 outputs)
+5.0 0.2 V (2 outputs)