Figure 2. Location of TI controls and instruments used during calibration.
(3) Adjust the AIR FLOW VALVE until the
8.
AIR FLOW Gage. a. Performance Check.
AIR FLOW gage (fig. 2) indicates 85. The flowmeter
(1) Connect the test equipment as shown in
(A1) must indicate between 225 and 235.
(4) Release the INHALE-EXHALE switch.
(2) Set and hold the INHALE-EXHALE switch
b. Adjustments. No adjustments can be made.
to EXHALE.
4