TB 55-4920-346-34
b. When all parameters are within tolerance, annotate and affix DA Label 80 (US Army Calibrated Instrument).
When the TI receives limited or special calibration, annotate and affix DA Label 163 (US Army Limited or Special
Calibration). When the TI cannot be adjusted within tolerance, repair the TI in accordance with the maintenance manual.
When repair is delayed for any reason or the TI cannot be repaired with local resources, annotate and affix DA Form
2417 (US Army Calibration System Rejected Instrument)and inform the owner/user accordingly in accordance with TB
750-25.